Topic: Measurements and Instrumentation
A type K thermocouple has sensitivity of 40µV/°C. The cold junction is maintained at 20°C and hot junction is at 320°C. Due to finite resistance of thermocouple wires, loading effect causes a loss in signal, resulting in an output voltage of 11.76 mV. If the measuring meter does not correct for this loading loss, what temperature will be displayed on meter?
- A. 314°C
- B. 300°C
- C. 294°C
- D. 274°C
Answer: A. 314°C
Explanation: The measuring meter calculates the temperature difference based strictly on the measured voltage and the ideal sensitivity. The measured voltage is $V_m = 11.76 \text{ mV} = 11760 \mu\text{V}$. The measured temperature difference is $\Delta T_{measured} = \frac{V_m}{\text{Sensitivity}} = \frac{11760}{40} = 294^{\circ}\text{C}$. Since the reference (cold) junction is known to be $20^{\circ}\text{C}$, the meter will calculate and display the hot junction temperature as $T_{display} = \Delta T_{measured} + T_{cold} = 294 + 20 = 314^{\circ}\text{C}$.